Main specifications
96
・
Specification of each measurement parameters
Measurement
parameter
Item
Description
pH
Measuring principle
Glass electrode
Display range
pH
2.00 to pH 16.00
Measuring range
pH 0.00 to pH 14.00
Resolution
0.01 pH
Accuracy
±0.01 pH
mV (ORP)
Measuring range
±2000 mV
Resolution
1 mV
Repeatability
1 mV or 1% of the read value,
whichever is greater
Temperature
Measuring principle
Thermistor method
Display range
30.0°C to 130.0°C
Measuring range
0.0°C to 100.0°C
Resolution
0.1°C
Accuracy
0.4°C
Dissolved
oxygen
Measuring principle
Membrane galvanic cell
Measuring range
(Display range)
0.00 mg/L to 20.00 mg/L
Resolution
0.01 mg/L
Accuracy
±0.1 mg/L
Oxygen
concentration
Measuring principle
Conversion from dissolved oxygen value
Measuring range
(Display range)
0.0% to 50.0%
Resolution
0.1%
Saturated
oxygen
concentration
Measuring principle
Conversion from dissolved oxygen value
Measuring range
(Display range)
0.0% to 200.0%
Resolution
0.1%