4 Measurement
28
4.4.2
Measurement condition setting (Meas. Set)
Q
Selecting the measurement range (Range)
Selects your target range.
1.
Referring to "
Q
Opening each setting menu" (page 23), display the measurement
condition setting screen “Meas. Set Menu”.
With this screen, measurement condition setting item can be selected.
2.
Select “Range” and press the ENT key.
The display flashes to show now the setting can be done.
3.
Press the
▲
/
▼
key to select the measurement range from 1000 ppb, 500.0 ppb and
100.0 ppb.
4.
Press the ENT key to confirm the setting.
Tip
When pressing the MENU key, the menu goes back to the previous menu.
Q
Setting the number of times for moving average (M.AVG Set)
1.
Referring to "
Q
Opening each setting menu" (page 23), display the measurement
condition setting screen “Meas. Set Menu”.
With this screen, the measurement condition setting item can be selected.
2.
Select “M.AVG Set” and press the ENT key.
The screen display “M.AVG Set Menu”. With this screen, number of times for moving
average can be set.
Tip
1 time can change the responsiveness for 0.5 seconds.
Number of times for TOC moving average (TOC)
Set the number of times for TOC value moving average.
1.
Perform the procedures 1. and 2. in "
Q
Setting the number of times for moving
average (M.AVG Set)" (page 28).
2.
Select “TOC” and press the ENT key.
The display flashes to show now the setting can be done.
3.
Press the
▲
/
▼
key to set the number of times for TOC moving average.
▲
key: Value increases by 1 time.
▼
key: Value decreases by 1 time.
4.
Press the ENT key to confirm the setting.
Tip
z
We recommend the number of times for TOC value moving average to 10 minutes.
z
When pressing the MENU key, the menu goes back to the previous menu.