HORIBA
Operation
NOx Converter Checker CC-100E
15/ 24
On the CC-100E
At the analyser
Status
Reading log
Press the UP key
once
The O3 gas stops,
NO and O2 flow into
the analyser.
Step (d)
Reading (d)
Press the UP key
once
O2 delivery is
stopped, only NO
flows into the
analyser.
Step (B)
Reading (B)
Press the UP key
once
The N2 gas flows
into the analyser.
Step (Z)
The step ends
automatically about
60 seconds later.
End
Tip: To scroll back through the steps, press the DOWN button each time.
Data entry form
The efficiency of the converter is calculated as follows:
Efficiency of the converter =
Test sheet: Converter efficiency: Analyser no., XXXXXXX
Date
Auditor
Efficiency
(in %)
Reading
Reading
before
and after
the check
(a)
(b)
(c)
(d)
(A)
(B)