24
9
Shake the electrode to remove any air bubbles remaining
on the ion selective membrane.
10
Perform the first calibration.
Wait for a stable reading and perform the first calibration as described in
the Meter Instruction Manual.
Do not touch the stir bar with the electrode. It may
damage the electrode.
Make sure that the rubber
stopper of the internal solution
fill port is open during
calibration.
If it is closed, the internal
solution does not ooze out very
much from the liquid junction.
As a result, the potential of the reference electrode will
become unstable and susceptible to stir.
It is recommended to place the electrode in the electrode stand.
If air bubbles remain on the ion selective
membrane, it may result in inaccurate
measurement.
Note
Note
Tip
Tip
Shake the
electrode
View video
View video
Note
Note
Summary of Contents for 6582S-10C
Page 1: ...Instruction Manual Combination type Potassium Ion Selective Electrode 6582S 10C ELECTRODE...
Page 18: ...8...
Page 52: ...42...
Page 62: ...52...
Page 79: ...69 Specifications External dimensions Unit mm 16 5 O D 18 85 150...