45
Maintenance
C
onsumables
Consumables and replacement periods
The consumables that need to be replaced and the replacement periods
are as follows. Replace each part at the appropriate timing to ensure
accurate measurement.
Ion electrode tip replacement guide
When the slope displayed on the meter is less than 80%
during calibration, clean the electrode (
electrode” (page 38)) and perform the calibration operation
again. If the slope is still less than 80% after that, replace the
ion electrode tip.
The main unit of the electrode and ion electrode tip are not
repairable parts.
Service life
Replace after approx. 400 hours
Ion electrode tip
Replace consumables at the period whichever occurs first; approx. one
year after purchase or after approx. 400 hours of use, depending on sample
solution, measurement frequency, and measurement environment.
approx. 1 year
approx. 1 year
approx. 1 year
Hours of use
Replace
Tip
Tip
1 year
1 year
approx. 2 years
approx. 2 years
1 year
Service life
Hours of use
approx. 2 years
Replace
Main unit of
electrode
Replace after approx. 800 hours
Replace consumables at the period whichever occurs first; approx. two
years after purchase or after approx. 800 hours of use, depending on sample
solution, measurement frequency, and measurement environment.