DEVICE DESCRIPTION
2
11
VERSAFLOW VORTEX 200
www.honeywellprocess.com
34-VF-25-137 iss.1 GLO July 2019 US
The permitted ambient temperature range is indicated on the nameplate; depending on the
device version it is T
amb
= -40...+65
°
C / -40...+149
°
F.
The minimum product temperature is -40
°
C / -40
°
F.
When using in zone 21, observe the maximum permissible product and ambient temperatures
for temperature class 2 and the information for use of a heat-resistant connecting cable and
cable entry.
Max. permissible product and ambient temperatures with signal converter or
connection box mounted above the flow sensor
Figure 2-3: Signal converter above the flow sensor
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
C
60
65
60
65
60
65
40
60
65
40
60
65
Nominal size
DN15
…
25
80
65
100
100
135 135
1
200 200
1
165
1
240
200
1
165
1
DN40
…
50
80
65
100
100
135 135
1
200 175
1
150
1
240
175
1
150
1
DN65
…
100
80
65
100 100
1
135
1
130
1
200 150
1
130
1
235
1
150
1
130
1
DN150...300
75
65
100
100
135 135
1
200 185
1
155
1
240
185
1
155
1
Table 2-7: Temperature class in
°
C
1
Permanent service temperature of connecting cable and cable entry min. 80
°
C
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
F
140 149 140
149
140
149
104
140
149
104
140
149
Nominal size
DN15
…
25
176
149
212
212
275 275
1
392 392
1
329
1
464
392
1
329
1
DN40
…
50
176
149
212
212
275 275
1
392 347
1
302
1
464
347
1
302
1
DN65
…
100
176
149
212 212
1
275
1
266
1
392 302
1
266
1
455
1
302
1
266
1
DN150...300
167
149
212
212
275 275
1
392 365
1
311
1
464
365
1
311
1
Table 2-8: Temperature class in
°
F
1
Permanent service temperature of connecting cable and cable entry min. 176
°
F