2
DEVICE DESCRIPTION
14
VERSAFLOW VORTEX 200
www.honeywellprocess.com
34-VF-25-137 iss.1 GLO July 2019 US
Max. permissible product and ambient temperatures for devices with painted flow
sensors / signal converters or connection box mounted at side or underneath the flow
sensor
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
C
60
65
60
65
60
65
40
60
65
40
60
65
Nominal size
DN15
…
25
65
65
95
90
120 120
1
120 120
120
1
120
120
120
1
DN40
…
50
65
65
85
80 120
1
120
1
120 120
1
120
1
120
120
1
120
1
DN65
…
100
65
65
95
90
1
120
1
120
1
120 120
1
120
1
120
120
1
120
1
DN150...300
65
65
85
85
120 120
1
120 120
120
1
120
120
120
1
Table 2-13: Temperature class in
°
C
1
Permanent service temperature of connecting cable and cable entry min. 80
°
C
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
F
140 149 140
149
140
149
104
140
149
104
140
149
Nominal size
DN15
…
25
149
149
203
194
248 248
1
248 248
248
1
248
248
248
1
DN40
…
50
149
149
185
176 248
1
248
1
248 248
1
248
1
248
248
1
248
1
DN65
…
100
149
149
203 248
1
248
1
248
1
248 248
1
248
1
248
248
1
248
1
DN150...300
149
149
185
248
248 248
1
248 248
248
1
248
248
248
1
Table 2-14: Temperature class in
°
F
1
Permanent service temperature of connecting cable and cable entry min. 176
°
F