1.
TMS 9250 Torque Measurement System
008-0738-00
10
TMS 9250 Product User Manual
Rev B
April 2012
© 2012 – Honeywell International Inc. All rights reserved
CAUTION
•
The SPM is intended only for connection to the specific power
supply shipped with the system.
•
The system was evaluated for use with a specific SELV power
supply of limited energy output that affords electrical fire and shock
hazard protection.
•
The use of another power supply with this system may result in
electrical shock or fire hazards or non-compliance of certification
standards.
•
The CCM is intended only for connection to the SPM with cables
provided by the manufacturer.
•
Refer to the APPENDIX F for connecting cables to the SPM.
•
This equipment provides only low voltage double insulated signal
I/O for connection to other equipment.
ATTENTION:
•
The product shall be in “POWER ON” condition for minimum period
of 30-45 minutes for meeting the product output specifications and
emission standards.
ATTENTION:
•
Ensure that the interface cables are shielded before connecting to
the product.
•
Due to the multitude of variations possible for connection of the
torque sensor to the load and prime mover, you must increase the
guarding for the end application to meet provisions of the
machinery directive.
•
The equipment is evaluated only with respect to the Low Voltage
Directive and RTTE Directives and is not evaluated to the
machinery directive for the reasons stated. Therefore, you must
ensure the compliance of this equipment with the provisions of the
machinery directive in the end application. This will normally
mandate the need for additional guarding at the mechanical
coupling points to the torque sensor.
TIP
•
Place the power supply of the system in a place where you can
easily unplug the power cords.