50
SPM Flex Gas Detector
Instrument Fault/Maintenance Fault/Information Codes
Type Sub-
type
Display
String
Technical
Description
Meaning of
Parameter
Probable
Cause
Corrective
Action
Instrument Faults
101
Flow Failure
Flow failure
Flow, cc/min
Pump failure
-Flow line obstruction
-Positive pressure
Replace pump
Check flow lines
Check for positive
pressure
103
Gate Motor Failure
Gate timeout
N/A
Gate motor failure
-Gate motor sensor
failure
Replace
104
Tape Advance Failure
Tape advance timeout
N/A
Broken tape
-Tape drive motor
failure
-Encoder failure
Replace
105
High Background
optics counts too low after multiple pulls
N/A
Tape issue
Check, replace
106
Gas Table Invalid
Gas table invalid
error code
CC RFID issue
Replace, contact HA
107
Comm� Failure RFID�
comm� fail to RFID reader
N/A
RFID PCB failure
Replace
108
Chemcassette Read Failure
RFID read failure
N/A
CC RFID issue
Replace, contact HA
109
Temperature Too High
overtemperature
Temperature, C
Environment is
over recommended
temperature
Improve ventilation/
cooling
110
Temperature Too Low
undertemperature
Temperature, C
Environment is
below recommended
temperature
Bring environment to
correct temperature
111
Optics Error
value of LED drive or reference detector signal
out of range
N/A
Optics PCB failure
Replace
112
Low Voltage Fault
DC input voltage low
voltage in Volts
Faulty power circuit
Contact HA
113
Chemcassette Expired
Now>LiMfg or
Now>LifeOFirstUse
1�0 or 2�0 for two causes
Chemcassette
cartridge has expired
Replace
114
Optics Subsystem Fail
no comm� to optics block
N/A
Optics cable
disconnected
-Optics PCB failure
Reconnect, replace
115
Comm� Failure GD
no comm� to GD
N/A
Main PCB failure
Contact HA
117
NV Memory Corrupt
CRC error
code to which test failed�
PCB failure
Replace
118
mA Output Mismatch
Feedback ADC mismatches DAC
Error magnitude in mA�
Faulty circuit
Contact HA
Incorrect external
wiring
check wiring
Bad calibration
Perform mA
calibration
119
Ethernet Fail
Ethernet hardware in UI failed to initialize
N/A
External connection
issue
Main PCB failure
Check external
Ethernet connection
Replace
120
Battery Very Low
Battery critically low, shutting down
Battery level
Low battery
Charge detector
Maintenance Faults
1
Flow Warning
Flow is unregulated
Flow, cc/min
Flow out of control
range
Check Chemcassette
Check sample tube
2
Chemcassette Low
Tape will run out in 24 hrs
N/A
Low tape
Replace
3
Idle Timeout
out of monitor too long
N/A
Out of monitor
Put in monitor and
reset fault
4
Low Voltage Warning
DC input voltage low
voltage in Volts
5
Comm Failure SPI
no SPI comm� to UI
N/A
Contact HA
6
File System Error
File system error in NAND or USB�
N/A
Contact HA
7
Comm� Invalid GD
Invalid SPI parameter data from GD�
N/A
Contact HA
8
Warn CC Expiring
A stored tape will expire in 7 days; if enabled,
the same fault will be issued for cartridges
in use�
N/A
Old tape
Replace
9
Software Error
catch all
error code
Contact HA
10
Optics Adjust Fail
Calibration failure
error code
Calibration failure
Perform calibration
again, Check the
calibration setup,
Contact HA
11
Flow Cal Fail
Calibration failure
error code
Calibration failure
Perform calibration
again, Check the
calibration setup,
Contact HA
12
Inhibit Timeout
Timed out
N/A
Detector has been left
in inhibit mode longer
than timeout warning
No action required
13
Force mA Timeout
Timed out
N/A
Detector has been in
force mode too long
No action required
14
Force Relay Timeout
Timed out
N/A
Detector has been in
force mode too long
No action required
15
Battery Low
Battery Low
Battery level
Battery is low
Charge battery
continued���
Summary of Contents for SPM Flex
Page 1: ...SPM Flex Single Point Monitor Gas Detector User Manual ...
Page 2: ......
Page 27: ...27 SPM Flex Gas Detector Setup Installation drawing Figure 18 Installation drawing ...
Page 70: ...70 SPM Flex Gas Detector ...
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ...SPM Flex Gas Detector User Manual 1998M0845 Revision 07 October 2016 2016 Honeywell Analytics ...
Page 77: ......