Performance Specifications
*Reproduces the transmitter signal exactly to within its resolution
Parameter
Deg. C Deg. F
Ambient Temperature
Reference
-23 +/-2 °C -73 +/- 2°F
Rated
-40 to 85 °C -40 to +185°F
Operative
-20 to 70 °C -4 to +158°F
Transportation/Storage
-50 to +100°C -58 to +212°F
Humidity (% RH)
Rack Mounting Field Housing
Reference
10 – 55% RH 10 – 55% RH
Rated
5 - 95% RH 5 – 100% RH
Operative
5 – 100% RH 5 – 100% RH
Transportation/Storage
5 – 100% RH 5 – 100% RH
Design Specifications
Input Signal
4-20 mA
Accuracy:
•
Analog 4-20 mA Mode
•
Temperature Effect
•
DE
+/- 0.1% of Span @ 23 +/-2 °C (+/- 1 digit)
+/- 0.01 % of Span /°C (+/- 1 digit)
Same as transmitter*
Honeywell Digital (DE) Mode
Reproduces the transmitter signal exactly to within its
resolution
Display
6 digits numeric indication
Display Resolution
0.01 unit for reading range (-999 to 999)
0.1 unit for reading range (-9999 to -1000) or (1000 unit to 9999).
1 unit for reading range (-99999 to -10000) or (10000 to 99999 ).
10 units for reading range (-999999 to -100000 ) or (100000 to
999999).
Display Update Rate
2 seconds
Display Indication Accuracy
0.1% of calibrated span (+/-1 digit)
Maximum Meter Voltage
30V DC
Max Loop Voltage Drop
≤
2.3V @ 21.5mA
Min Loop Operating Current
3.6mA
Power Fail
All calibration constants are stored in non-volatile memory
and are not affected by power loss.
Failure mode
LCD display failure will not affect transmitter operation
Reverse Polarity Protection
Yes
Loop Supply Voltage
Requirement
Min. loop voltage = 2.3Vdc + (Rloop * 22mA)
Where:
MinVTx is the transmitter minimum supply voltage
Rloop is the resistance on the loop (in ohms)
Display Performance
The LCD display will turn blank below -20 °C and above
70 °C
(-4°F and 158°F), rendering the display unreadable.
However the loop will continue to function.
Summary of Contents for SmartLine RMA801
Page 2: ......
Page 11: ......
Page 50: ...39 RMA803 Remote Indicator User s Guide Revision 2 ...
Page 51: ...Revision 2 RMA803 Remote Indicator User s Guide 40 ...
Page 52: ...41 RMA803 Remote Indicator User s Guide Revision 2 ...
Page 57: ...Revision 2 RMA803 Remote Indicator User s Guide 46 Control Drawing 50089981 ...
Page 58: ...47 RMA803 Remote Indicator User s Guide Revision 2 ...
Page 59: ...Revision 2 RMA803 Remote Indicator User s Guide 48 ...