Satellite XT R Version
5-1
Calibration - Section 5
This section describes calibration procedures for the Satellite XT.
Calibration can be performed either automatically using a dynamic gas calibration
or manually by entering a calculated correction factor called K-factor.
Main Menu - Calibration
The instrument is in the Maintenance Mode and the
green LED is off. Actuate the cursor keys <up> or
<down> to move forwards and backwards in the Main
Menu. To select a submenu, actuate the <set> key to
enter.
Sub menus - Calibration
Gas Calibration
screen 1 of 2
Manual K-Factor
screen 2 of 2
!Note:
If a high degree of accuracy in monitoring is required, monthly calibration with calibration gas
of a known concentration is recommended.
When performing a dynamic calibration, a new correction factor (K-factor) is calculated
automatically. The actual value can be displayed in the submenu Manual K-Factor.
Use appropriate safety precautions when handling toxic or corrosive gases and properly
vent, if possible.
! Instruments with OXYGEN sensors:
Span Adjustment must be performed prior to start-up. It is recommended to repeat Span
Adjustment every 4 to 6 weeks. For details refer to Section 5, Calibration.
CA L I BRA T I ON
s e l e c t :
3
o f
4
M E N U
s e l e c t :
1
o f
2
C A L I B R A T I O N
GA S CA L I BRA T I ON
s e l e c t :
2
o f
2
C A L I B R A T I O N
MANUA L K - F AC TOR