MeshGuard LEL IR & CO2 IR User’s Guide
32
6.3.
Diagnostic Mode Readings
In Diagnostic mode, you can step through readings by pressing [N/-].
6.3.1.
Sensor Raw Count
Sensor Raw Count is indicated by “rAU” followed by three dashes (- - -). This indicates that the
sensor has digital output that does not produce a raw count.
•
Press [N/-] to advance to the next reading.
•
Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.2.
Battery Raw Count
Battery Raw Count is indicated by “bAt” followed by a number.
•
Press [N/-] to advance to the next reading.
•
Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.3.
Temperature Raw Count
Temperature Raw Count is indicated by “tNp” followed by a number.
•
Press [N/-] to advance to the next reading.
•
Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.4.
Zero Raw Count
Zero Raw Count is indicated by “0rC” followed by three dashes
(- - -). This indicates that the sensor has digital output that does not produce a raw count.
•
Press [N/-] to advance to the next reading.
•
Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.
6.3.5.
Calibration Delta Counts
Calibration Delta CTS is indicated by “CdC” by three dashes (- - -). This indicates that the sensor
has digital output that does not produce a raw count.
•
Press [N/-] to return to the first raw count.
•
Press [MODE] and [Y/+] together for three seconds to exit Diagnostic Mode and enter
Detection Mode.