Introduction
1.4 Specifications
General:
Displayed Process
Variable
DL424: 0-20 ppm dissolved oxygen
DL425: 0-200 ppb dissolved oxygen
Displayed
Temperature Range
2.0 to 60.0 ºC (35.6 to 140ºF)
Operating
Temperature
–20 °C to +60 °C (-4 °F to +140 °F)
Storage Temperature
–20 °C to +70 °C (-4 °F to +158 °F)
Display Resolution
DL424: 0.01
ppm
DL425: 0.1 ppb in the 0-20 ppb range; 1 ppb in the
0-200 ppb range
Output Type
4-20 mA (2-wire loop powered)
Output Scale
DL424: 0-20ppm
DL425: 0-200ppb
Output Calibration
4-20 mA
Power
16-42 Vdc
Maximum Load Resistance:
250 ohms at 16 Vdc
600 ohms at 24 Vdc
1400 ohms at 42 Vdc
Calibration Options
Air Cal; Sample Cal
Diagnostics
Sensor and electronics
Physical Properties:
Case
Weatherproof, corrosion-resistant plastic housing
Local Display and
Buttons
LCD 4-digit, 7-segment
Field Wiring Cordset
Shielded twisted pair. Length: 6 m (19.7’)
User Termination
Tinned leads
Remote Mounting
Pipe, Wall, or DIN Rail
Dimensions
H 123 mm (4.84”) x W 48 mm (1.89”) x D 46 mm (1.81”)
Weight
Approximately 142 g (5.0 oz.)
Approvals
CE Mark - Industrial Applications
UL – General Purpose
CSA – General Purpose
IP66 Enclosure
NEMA Type 4
FM Class1, Div. 1 (I.S.)
FM Class 1, Div. 2 (non-incendive field wiring)
9/11 DirectLine®
DL424/425 Sensor Module User Manual
3