5
The Sampler
Operating Guidelines
Standard Hose Configuration
Sample Tube Connections
The Sampler is shipped with the short sample tube installed.
To remove the installed sample tube, grasp the end of the tube and pull straight out. Insert the
new sample tube through the enclosure and into the gasket opening. To ensure a proper seal,
the end of the tubing should be inserted until firm resistance is felt.
Block the Airflow
Block the airflow to the Sampler to confirm that it is working correctly and there are no leaks in
the system.
There are two ways for you to test the airflow of the Sampler:
1. Press and hold the block trigger
2. Kink the sampling or attachment hose.
If the red
FAIL
LED blinks and the alarm beeps while the airflow is blocked, then the Sampler is
operating correctly. If it does not, reconnect the hose/tube to check for leaks.
Note:
We recommend that you perform a block test once the
Sampler is activated.
Sampling hose
Attachment hose