CAUTION
Move to a normal atmosphere (20.9% v/v O
2
) that is free of hazardous gas. Use 50% LEL for
test gas.
Details for calibration and maintenance:
l
Recommendations for calibration of the equipment on a routine basis including the
maximum time interval between calibrations.
l
Calibrate the apparatus before first-time use and then on a regular schedule, depending on
use and sensor exposure to poisons and contaminants. Honeywell recommends that the
sensors be calibrated regularly and at least once every 180 days (6 months).
l
The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different
combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
Procedure to calibrate the detector via the calibration cap and the DC app on a
mobile device
1.
Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs
to snap it into place.
BW Icon
11
User Manual