background image

AQS51/61-KAM  COMBINED CO

2

-/TEMPERATURE DUCT SENSOR/CONTROLLER

5

EN0B-1023 GE02R0700

Fig. 3  Mounting of AQS51/61-KAM onto the duct

Mounting Instructions

1. Drill a 

25 mm hole for the sampling probe (Fig. 2,

Pos.1) into the duct of the selected mounting area.

2. Use the gasket (Pos. 2) to mark on the duct surface the

positions of the two holes for the mounting screws.

3. Drill 

4mm holes for the mounting screws (Pos. 5).

4. Insert the sampling probe (with the gasket) into the duct.

Position the probe relative to the duct so that the air inlet is

on the left hand side and faces toward the duct air flow.

For flow direction see also the arrow on the flow direction

label (Pos. 3).

NOTE:

In cases where the sampling probe can’t be

mounted with air flow direction from the left to

the right, reversed flow direction through the

sampling probe is possible, but influences the

response time slightly. Use the enclosed arrow

sticker to indicate on the probe the opposite and

correct air flow direction

5. Fasten the sampling probe and the sealing gasket onto the

duct with the two screws (Pos. 5).

6. Remove the cover of the aspiration box. The box has

several knock outs in two dimensions for PG7 (Pos. 8)

and PG9 (Pos. 7) cable entry bushings, To make a hole,

insert a screwdriver into the groove and hit firmly. Mount

for each cable a cable entry bushing into the aspiration

box and seal properly.

7. Mounting the aspiration box to the sampling probe is

performed by a snap-in bayonet fitting (Pos. 10). Insert first

the temperature sensor placed on the end of a 150mm

long soldered-on lead (Fig. 3, Pos. 2) into the air inlet to

ensure accurate temperature measurement. Turn the box

by approx. 23

°

 ccw and stick the box onto the sampling

probe, so that the probe is fitted into the notches of the

box, then turn the box cw until stop (see Fig. 2).

8. Feed the cables through the cable entry bushings and seal

the cable entries carefully by turning the nut to prevent

ambient air suction into the aspiration box during

operation. Feed never more than one cable through a

cable entry bushing.

  9. Select the required output ranges by setting the three

coding jumpers accordingly to Table 1. The default

settings are for 0...10Vdc outputs as indicated by the

darker jumper positions in Fig. 5. The outputs have to be

configured before the unit is connected to the power

supply.

Do not change jumper positions when power is on.

10. Connect the cable wires to the terminal block as listed

in Table 2.

11. Mount the cover.

Fig. 4  Printed Circuit Board

Fig. 5  Jumper positions

1

2

1 Hole with 

25mm

2 Temperature sensor with 150mm cable

Summary of Contents for AQS51-KAM

Page 1: ...s analog output configuration the software program STRATEGY is available which is part of the User Interface Program Software package AQS USP22 used for re calibration change of the measurement range or other service or set up purposes The switching setpoint and differential of the relay output for CO2 limit monitoring or plant system ON OFF control can be adjusted within the measurement range Thi...

Page 2: ...D indication Yellow Maintenance required Red Relay activated Power up time 1 min Ambient operating limits Temperature 0 50 C General Humidity 0 95 rh non condensing Operating principle Non dispersive infrared NDIR Gas sampling mode Diffusion and 3 mm gas inlet push on barb Response time 2 min diffusion or 10 sec 0 2 ltr min gas flow Measurement range 0 3000 ppm Converting range 0 2000 ppm 0 10 Vdc...

Page 3: ...using the PC software STRATEGY Version 3 Digital relay output Single pole double throw SPDT switching potential free contact Factory adjusted switching setpoint and differential 1A 50 Vac 50VA or 1 A 24 Vdc 30 W Relay activated 1000 ppm Relay deactivated 900 ppm Sensor housing 84 x 142 x 46 mm Dimensions Sampling probe 203 mm Zigzag traces accessible after removing the cover Shortage with a screwd...

Page 4: ...ions to provide different ventilation control strategies All sensor measurements can be addressed to these functions and mixed together by this software to form two independent and intelligent P I controllers The relay output contact CM NO is closed if the CO2 level exceeds the factory adjusted limit of 1000ppm and the contact CM NC is closed if the CO2 level is equal or below 900ppm These paramet...

Page 5: ...Pos 7 cable entry bushings To make a hole insert a screwdriver into the groove and hit firmly Mount for each cable a cable entry bushing into the aspiration box and seal properly 7 Mounting the aspiration box to the sampling probe is performed by a snap in bayonet fitting Pos 10 Insert first the temperature sensor placed on the end of a 150mm long soldered on lead Fig 3 Pos 2 into the air inlet to...

Page 6: ...G0 permitting a 3 wire configuration A single transformer may be used for the entire system Situations can occur where different transformers must be used NOTE2 AQS51 61 can supply both a voltage or a current output for OUT1 OUT2 To change between voltage and current output mode the hardware jumpers have to be set in accordance There is one jumper for AN1 and one for AN2 so that one output can be ...

Page 7: ...2 Error code 3 CO2 concentration in ppm 4 Temperature in C 5 OUT1 value full scale 6 OUT2 value full scale To access the maintenance level short at level 5 OUT1 value ENTER MENU ENTER in sequence The maintenance mode is restricted to competent trained service personnel only For detailed information see the Service Checkout Instruction EN2B 0461GE02 LED Signals The standard sensor has two different...

Page 8: ...sor probes are checked against failure by checking the valid dynamic measurement ranges Error Code and Actions Error Code Suggested Action 4 Indicate problems with the power supply Check power supply voltage 8 Indicates problem with OUT1 Could be too high an output load connected to the output Check OUT1 with a voltmeter and compare the output with the LCD reading for AQS61 KAM Disconnect the outp...

Reviews: