Introduction
Page 6
1
Introduction
The Honeywell Enraf 854 ATG (Advanced Technology Gauge) measures the liquid level and can be
programmed to measure two additional interface levels.
The 854 ATG has four programmable level alarms, and also provides diagnostic information. This information
can be displayed on the internal display, the Portable Enraf Terminal (PET) as well as on remote systems.
Optionally, the instrument can be provided with software for density measurement of the stored product.
The 854 ATG can be supplied with an optional board (called: HCU board) to interface optional equipment as:
•
Spot temperature element (Pt100 RTD)
•
Average temperature element (VITO MTT/LT temperature probe, MRT element, MPT element)
•
Water bottom probe (VITO water probe, VITO combi probe or external HART water probe)
•
Pressure transmitters (for vapour pressure measurement, HIMS density measurement)
•
Analog 4-20 mA level output
The above functions are separately available or in combinations as with the previous MPU, HPU, HSU and
TPU optional boards. Refer to the below table:
Sales code
option board
(4
th
position)
Available functions
HCU board
Emulation
mode
B
Spot temperature Pt100 RTD
TPU-2/ HSU
C
VITO temperature and/or water probe
HPU
J
VITO temperature and/or water probe + HART device(s)
HPU
U
Spot temperature Pt100 RTD + HART device(s)
HSU
V
Analog level output
MPU
W
Analog level VITO temperature and/or water
probe
HCU
X
Analog level VITO temperature probe
MPU
Y
Analog level Spot temperature Pt100 RTD +
VITO temperature and/or water probe + HART device(s)
HCU
Notes:
1 HART device(s) can be:
•
HART pressure transmitters for vapour pressure measurement or HIMS density measurement
•
HART water bottom probes.
2 With sales code Y (all HCU functions), the spot temperature measurement
is disabled if a VITO temperature probe is present.
Summary of Contents for 854 ATG Series
Page 1: ...Instruction manual Series 854 ATG level gauge ...
Page 2: ......
Page 53: ...Appendix Instruction manual 854 ATG Page 51 ...
Page 55: ...Appendix Instruction manual 854 ATG Page 53 ...
Page 56: ...Appendix Page 54 Appendix C ASCII table ...
Page 58: ...Appendix Page 56 Level display format B Average gas temperature display format C ...
Page 60: ...Appendix Page 58 HIMS pressure displays format F and H Servo density display format I ...
Page 62: ...Appendix Page 60 Appendix E PCB layout XPU 2 HCU HCU SPU II GFC ...
Page 67: ......