Maintenance
Instruction manual 854 ATG
Page 45
Do not yet reprogram this value in the gauge. Reprogram all values after completion of all the four calibration
measurements. Repeat this calibration from item
BT
for the other 3 test weight combinations:
•
for
F1
with: (25+75 g),
•
for
F2
with: (25+150 g),
•
for
F3
with: (25+225 g).
Remove the test weights from the measuring wire and attach the displacer.
Program the measured frequencies :
Item
Name
Description
W2
=
Protection level 2
Enter protection level 2 password.
F0
= Frequency constant 0
Standard floating point format; unit: Hz. Program frequency 0,
measured with test weight of 25 g.
F1
= Frequency constant 1
Standard floating point format; unit: Hz. Program frequency 1,
measured with test weight of 100 g.
F2
= Frequency constant 2
Standard floating point format: unit: Hz. Program frequency 2,
measured with test weight of 175 g.
F3
= Frequency constant 3
Standard floating point format; unit: Hz. Program frequency 3,
measured with test weight of 250 g.
WT
= Wire rupture
Three characters. Reprogram wire tension protection
(advised setting: EDE).
EX
Exit
Exit protection level.
Note:
A faster method of calibration is using the commands
MF
and
FQ
in stead of
BT
and
BF
. This method
may not be followed when your gauge is used for density measurement via the density displacer.
5.6 Synchronizing the reference encoder
After the installation of a new motor block or in case of mounting new software, the internal reference
encoder must be synchronized to the position of the reference encoder and the gauge starts with an error
code e.g. ES553 / ES555. The following procedure will do.
Item
Name
Description
W2
=
Protection level 2
Enter protection level 2 password.
SM
Set maintenance
Go into maintenance mode and do not enter any command which is
not specified below.
FP
Find position
The 854 ATG finds its encoder position. Wait appr. 20 seconds.
SO
Set operational
Restart in operational mode.
EX
Exit
Exit protection level and go back to operational mode.
The gauge is now ready for level calibration (refer to section 3.4).
Summary of Contents for 854 ATG Series
Page 1: ...Instruction manual Series 854 ATG level gauge ...
Page 2: ......
Page 53: ...Appendix Instruction manual 854 ATG Page 51 ...
Page 55: ...Appendix Instruction manual 854 ATG Page 53 ...
Page 56: ...Appendix Page 54 Appendix C ASCII table ...
Page 58: ...Appendix Page 56 Level display format B Average gas temperature display format C ...
Page 60: ...Appendix Page 58 HIMS pressure displays format F and H Servo density display format I ...
Page 62: ...Appendix Page 60 Appendix E PCB layout XPU 2 HCU HCU SPU II GFC ...
Page 67: ......