PRODUCT DESCRIPTION
Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate
can be measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate.
The most recent 60 measurements can be stored in the memory with date and time stamp. The Electronic
Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN 60601-1:2006/AC:2010
(Medical electrical equipment – Part 1: General requirements for basic safety and essential performance),
IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
- Requirements and tests), IEC 80601-2-30:2009+Cor.2010/EN 80601-2-30:2010 (Medical electrical equipment
– Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive
sphygmomanometers),EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1:
General requirements), EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part
3: Supplementary requirements for electro-mechanical blood pressure measuring systems). ANSI/AAMI
SP-10:2002+A1:2003+A2:2006
NOTICE
1. Read all of the information in the operation guide and any other literature in the box before operating the unit.
2. Stay still, calm and rest for 5 minutes before blood pressure measurement.
3. The cuff should be placed at the same level as your heart.
4. During measurement, neither speak nor move your body and arm.
5. Measuring on same wrist for each measurement.
6. Please always relax at least 1 or 1.5 minutes between measurements to allow the blood circulation in your
arm to recover. Prolonged over-inflation (cuff pressure exceed 300 mmHg or maintained above15 mmHg
for longer than 3 minutes) of the bladder may cause ecchymoma of your wrist.
7. Consult your physician if you have any doubt about below cases:
1) The application of the cuff over a wound or inflammation diseases;
2) The application of the cuff on any limb where intravascular access or therapy, or an arterio-venous (A-V)
shunt, is present;
3) The application of the cuff on the arm on the side of a mastectomy;
4) Simultaneously used with other monitoring medical equipments on the same limb;
5) Need to check the blood circulation of the user.
8. This Electronic Sphygmomanometers is designed for adults and should never be used on infants or
young children. Consult your physician or other health care professionals before use on older children.
9. Do not use this unit in a moving vehicle, This may result in erroneous measurement.
10. Blood pressure measurements determined by this monitor are equivalent to those obtained by a trained
observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American
National Standard Institute, Electronic or automated sphygmomanometers.
11. Information regarding potential electromagnetic or other interference between the blood pressure
monitor and other devices together with advice regarding avoidance of such interference please see part
ELECTROMAGNETIC COMPATIBILITY INFORMATION.
12. If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood
pressure measurement, a signal of
will be displayed. Under this condition, the Electronic
Sphygmomanometers can keep function, but the results may not be accurate, it’s suggested that you
consult with your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s, and the number of such pulse takes more than 53
percentage of the total number of pulse.
13. Please do not use the cuff other than supplied by the manufacturer, otherwise it may bring biocompatible
hazard and might result in measurement error.
SPECIFICATIONS
Product name:
Blood Pressure Monitor
Model:
BPW-1005
Classification:
Internally powered, Type BF applied part, IPX0, No AP or APG, Continuous operation
Machine size:
Approx. 85mm x 64.5mm x 28mm (3 11/32”x 2 17/32”x 1 3/32”)
Accuracy:
Pressure: 5ºC-40ºc within±0.4kpa (3mmHg) pulse value:±5%
Cuff circumference:
14cm - 19.5cm (5 1/2” - 7 11/16”)
Weight:
Approx. 110g (3 7/8 oz.) (exclude batteries)
Measuring method:
Oscillometric method, automatic inflation and measurement
Memory volume:
60 times with time and date stamp
Power source:
Batteries: 2 × 1.5V
Size AAA
Measurement range:
Cuff pressure: 0-300mmHg
Systolic: 60-260mmHg | Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
Accuracy:
Pressure: ±3mmHg | Pulse rate: ±5%
Environmental temperature for operation:
5ºC~40ºC (41ºF~104ºF)
Environmental humidity for operation:
≤90%RH
Environmental temperature for storage and transport:
-20ºC~55ºC (-4ºC~131ºC)
Environmental humidity for storage and transport:
≤90%RH
Environmental pressure:
0kPa-105kPa
Battery life:
Approx 270 times
All components belonging to the pressure measuring system, including accessories:
Pump, Valve, LCD, Cuff, Sensor
Note:
These specifications are subject to change without notice.