Model LBP30
Smart
Pulsewave Blood Pressure Monitor
Table 1:
1
Guidance and manufacturer’s declaration – electromagnetic emission
2
The
Smart
Pulsewave Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of
Smart
Pulsewave Blood Pressure Monitor should assure that it
is used in such an environment.
3
Emissions test
Compliance
Electromagnetic environment - guidance
4
RF emissions
EN 55011
Group 1
The
Smart
Pulsewave Blood Pressure Monitor uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment
5
RF emissions
EN 55011
Class B
The
Smart
Pulsewave Blood Pressure
monitor is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
6
Harmonic emissions
EN 61000-3-2
Class A
7
Voltage fluctuations /
flicker emissions
EN 61000-3-3
Complies
Table 2:
Guidance and manufacturer’s declaration – electromagnetic immunity
The
Smart
Pulsewave Blood Pressure Monitor is intended for use in the electromagnetic environment specified
below. The customer or the user of the
Smart
Pulsewave Blood Pressure Monitor should assure that it is used in
such an environment.
Immunity test
EN 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with
synthetic material, the relative
humidity should be at least 30 %.