2.6. System Inputs and Outputs
In the preceding table, ‘Display Unit’ refers to a CheckMate Ultra display unit or
Mammolux motorized viewer. These devices (and the R2 postscript printer) can
display (or print) no more than four views per case.
If your system has been configured with the ImageChecker license, the R2 processing
unit can produce ImageChecker CAD results for all supported views. However,
ImageChecker results for certain views may not be available with some output devices
and formats.
View Modifiers
View modifiers can be added to describe any supported view. The following table lists
the supported DICOM view modifiers.
View Modifier
Label
View Modified
Axillary Tail
AT
MLO
Cleavage CV
CC
Partial View*
(none)
Any
Rolled Lateral
…RL
Any
Rolled Medial
…RM
Any
Rolled Inferior
…RI
Any
Rolled Superior
…RS
Any
Magnification M…
Any
Spot Compression
S
Any
Implant Displaced**
ID
Any
Implant Present*
(none)
Any
Tangential
TAN
Any
*The Partial View and Implant Present modifier labels are NOT added to the
view description, but are present in the DICOM header of the DigitalNow image.
** Implant-displaced views with a maximum of 2.5 cm (1 in) of implant imaged.
Important!
Certain view modifiers prevent the system from performing CAD
processing. For more information, see the manual,
Understanding ImageChecker 8.5–
8.7
.
R2 DM User Manual – PN MAN-01044 Rev 001
21
Summary of Contents for R2 DM
Page 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Page 2: ......
Page 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Page 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Page 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Page 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Page 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Page 105: ......
Page 106: ......