User Manual
Version 1.0 19
X.
Faults Diagnosis
Problem
Possible Cause
Solution
No Power
The power plug is not plugged in or
power not switched ‘
Check the power plug is
inserted, and power switched
‘ON
Fuse has failed.
Replace the fuse.
Cabinet Temperature
Does Not Increase
The set temperature is low.
Adjust the set temperature.
The electric heater is not working.
Replace the electric heater.
The temperature control instrument
is not working
Replace the temperature
control instrument.
The cyclic fan is not working
Replace the fan.
Set Temperature
Variance to the
Cabinet Temperature
The temperature sensor is not
working
Change the temperature
sensor.
The fine-tuning potentiometer for
setting temperature appears not
well adjusted.
Adjust the potentiometer.
Over-Temperature
Alarm is Abnormal
The set temperature is low.
Adjust the set temperature.
The temperature control instrument
is not working.
Replace the temperature
control instrument.