User Manual
Version 1.0 15
8.6 Start / Stop Operation
1
.
After setting the operating parameters, press
button for (about 4-5 seconds) to
start the heating cycle/cooling process with defined time in hh:mm or continue the
process for (00:00). When the word “END” disappears on the right corner of the screen,
the message “RUN” will appear in the left bottom and the system will display issues like
contemporary, timer, temperature measured inside the chamber.
2
.
At any time, you can always manually stop the cycle by pressing the
button for (4-
5 seconds).
3
.
Once the timer runs out or after a manual stop, the instrument beeps intermittently
and the icon of the visual alarm and the word
"end"
appears on the display. Press any
button to silence.
NOTE:
the acoustic signal will not end until it is stopped by the operator, but the heating
cycle will be terminated so the samples inside the instrument will remain exposed to the
internal temperature of the chamber.
8.7 Delay of Heating Cycle Start
1. It is possible to set a delay (hour and minutes) of the heating cycle start.
Simultaneously Press
and
for a
few seconds until the time position shows “LK
0000” confirm the “0000” password pressing one time
. On the top right part of
display the
parameter “dy” (delay) appears close to value 00:00.
2. Set the desired delay value (hh:mm) by pressing
or
keys. For quick
movement between the digits use the
button. Confirm the set value with another
press of
button. The display comes back to the standby screen.
3. Pressing the
button with long pressure (4-5 seconds) the instrument starts the
work cycle,
but it doesn’t immediately heat: the word “end” and the set delay time
alternately blink on the top right part of display, counting the wait time until the real
starting of heating. Once the delayed time is passed the instrument starts to heat and
the regular timer appears on the display.