If the probe is turned on during the fast charge period of

the  charge  cycle,  the  current  surge  that  the  battery

provides  to  the  electronics  in  the  probe  will  cause  a

momentary voltage drop on the battery, causing the

charger  to  terminate  the  fast  charge  based  on  a  peak

voltage  detect  condition. When a fast charging peak

voltage is detected, the charger COMPLETE LED (green)

will light.

The charger switches to Top-Off charge mode which

provides a charge current one eighth of the fast charge

rate.  This charge continues from the time when a peak

voltage detect terminates the fast charge, until the

maximum time of 180 minutes is reached. If the probe is

disconnected from the charger at this point in the charge

cycle, the battery will still operate properly.

Following  the maximum time of 180 minutes, a Pulse

trickle charge mode is used to compensate for the self

discharge of the battery while idle in the charger.

The batteries are quite small, and the fast charge current

is very low.  The fast charge current is low enough that

there is not always enough heating of the battery cells to

cause the voltage to drop.  With no voltage drop there is

no  peak  voltage  detection,  and  the  fast  charge will

terminate  after the maximum time of 180 minutes is

reached.

The chargers will not damage the battery even when left

connected for long periods of time.

Summary of Contents for 491198-48 Series

Page 1: ...Series 491198 48 NiMH Battery Fast Charger User s Manual Copyright 1999 2000 by Holaday Industries Inc Manual 600087 5 00 12 50...

Page 2: ......

Page 3: ...Revision Record Manual 600087 Series 491198 48 NiMH Battery Fast Charger Revision Description Date Release 4 99 A Changed Area Code 2 00 B Added charging information 5 00...

Page 4: ......

Page 5: ...rument dangerous Intentional interruption is prohibited DO NOT defeat the earth grounding protection by using an extension cable power cable or autotransformer without a protective ground conductor Th...

Page 6: ...and relays that are likely to produce sparks or arcs Therefore when charging a battery in a garage or in a similar location locate the battery charger in a room or enclosure provided for this purpose...

Page 7: ...REPAIRS WILL BE BILLED AT A NOMINAL COST IN SUCH CASE AN ESTIMATE WILL BE PROVIDED BEFORE WORK IS STARTED IF REQUESTED BY THE PURCHASER FOR WARRANTY SERVICE CONTACT HOLADAY INDUSTRIES INC GIVING FULL...

Page 8: ......

Page 9: ...N 1 2 0 UNPACKING AND ACCEPTANCE 3 Introduction 3 3 0 SPECIFICATIONS 5 4 0 OPERATING INSTRUCTIONS 7 Input Power Requirements 7 Fuse Replacement 7 Charging a Battery 9 Charging Indicators 9 Discharging...

Page 10: ......

Page 11: ...state of the battery has no effect other than shortening the charge time Housed in a rugged enclosure power enters the battery charger through a power entry module which contains the fuses or an optio...

Page 12: ...2 Page Fast Charger Manual Figure 1 4 8V Battery Charger...

Page 13: ...oncealed damage or loss retain the packing materials for inspection by the carrier Step 2 Remove the product from its shipping container Save boxes and any protective packing materials for future use...

Page 14: ...4 Page Fast Charger Manual...

Page 15: ...25 mA Pulsed Trickle Charge Current 2 0 mA Output Voltage During Fast Charge 4 8 Vdc NiMH Battery 4 8 Volt 4 Cell NiMH Battery 25 mAH Rapid charge cells 1 2 volts cell Holaday Part 491193 Environment...

Page 16: ...EE 7 7 UC 814 Molded PVC Receptacle IEC320 C13 UC 051 1 0mm2 3 Cond 10A 250V Countries Argentina Austria Brazil Finland France Germany Israel Italy Holland Korea Netherlands Norway Sweden Turkey 22260...

Page 17: ...use Fuse Replacement If a fuse has blown it must be replaced with the same type and value or an unsafe condition may result Refer to Figure 2 for fuse replacement The fuse is contained in the fuse dra...

Page 18: ...8 Page Fast Charger Manual Figure 2 Fuse Replacement...

Page 19: ...ct a battery the NO BATTERY LED amber will light When the charger does detect the HI 6005 s battery the PENDING LED amber lights Before fast charging can begin the battery voltage must fall within pre...

Page 20: ...terminates the fast charge until the maximum time of 180 minutes is reached If the probe is disconnected from the charger at this point in the charge cycle the battery will still operate properly Fol...

Page 21: ...DISCHARGE LED amber will light You may press the Discharge button at any time The battery is now being fully discharged a process that may take up to 5 hours When this cycle is complete it will then...

Page 22: ...12 Page Fast Charger Manual...

Page 23: ...wn fuse inside the power entry module refer to figure 2 CAUTION NEVER attempt to change the fuse with the battery charger plugged in If a fuse is blown it must be replaced Be sure to use only 250 Volt...

Page 24: ...14 Page Fast Charger Manual Notes...

Reviews: