20
Respiration Therapy Device
TREND 300
Functions of
TREND 300
WAKE TIME.......... 06:00
EXIT MENU!
Press Input key
to leave the programming mode.
During standard operation the display indicates
date and time:
S A 2 8 . F E B
1 2 : 4 2
The alarm clock can be activated or deactivated by the
keys
and
.
With the key
the alarm clock can be switched on. For
controlling the programmed alarm time is indicated for a short
time in the display. After this a bell symbol shows that
the alarm clock is active.
SA 28.FEB
12:42
With the key
the alarm clock can be switched off (e.g. on
Sunday). In this case the bell symbol is not indicated.
NOTE:
If there is no key pressed during programming for
30 seconds the programming mode will be left
automatically.
All functions are programmed in the same way:
Press
for 1 s, and choose the line requested with the
keys
and
:
By pressing the key
the following options can be
chosen:
WAKE TIME
WAKE HOUR, WAKE MINUTE
TIME
HOUR, MINUTE
DATE
YEAR, MONTH, DAY, WEEKDAY
RAMP
1.
mask test time: 0...90 s
(in 5 s-steps)
2. softstart time: 0...60 min
(in 1 min-steps)
AUTOMATIC
ON, OFF
I-SET
1.
PIN INPUT!
following selection after:
2. 3,0 ... 18,0 hPa
(in 0,5 hPa-steps)
E-SET
3,0 ... 18,0 hPa
(in 0,5 hPa-steps)
I-SLOPE
0,0 ... 1,5 s
(in 0,1 s-steps)
E-SLOPE
0,0 ... 1,5 s
(in 0,1 s-steps)
TRIGGER
1 ... 5
MODE
BILEV-S
(only indicate, not selection possibility)
BRIGHTNESS
100%, 50%
COUNTER
POWER, BLOWER, FILTER, THERAPY
(time indicate in h)
EXIT MENU
!
If you want to finish the programming, choose the menu
line
“
EXIT MENU
!
”
and press key
.