IM_POJM180417 ARM BLOOD PRESSURE MONITORBASIC 2018-07-25
TECHNICAL ALARM DESCRIPTION
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with
no delay if the determined blood pressure (systolic or diastolic)
is outside the rated range specified in part SPECIFICACIONS.
In this case, you should consult a physician or check if your
operation violated the instructions.
The technical alarm condition (outside the rated range) is preset
in the factory and cannot be adjusted or inactivated. This alarm
condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The
signal displayed on LCD will disappear automatically after
about 8 seconds.