RS -
0…32767
Read Status
Status word (see below)
RT -
0…100;
0…100;
0…100
Read Temperatures
VT_PA1 [0.1°C]: Temperature of output stage pos. branch
VT_PA2 [0.1°C]: Temperature of output stage neg. branch
VT_DC [0.1°C]:
Temperature of DC/DC converter
RV -
-3000…3000
Read Output Voltage
VOAV [V]:
Actual value of output voltage
SD
0,1
-
Select Setpoint Source
0: analog setpoint VONVA (default)
1: digital setpoint VONVD
SQ -
-
Shutdown
Request
Set Shutdown Latch
SR -
-
Shutdown
Reset
Reset Shutdown Latch
SS
0,1
-
Select Analog Input Source Resistance
0: 0
Ω
signal source output impedance
1: 50
Ω
signal source output impedance (default)
SV -3000…
3000
Set Output Voltage Setpoint
VONVD [10mV]: digital setpoint; resolution 10mV
Status word:
Bit # Status bit
Function
0
AXPWFAIL
Aux Power Fail: Failure of auxiliary su15V, -15V, +5.2V
1
IL
Interlock loop open
2
IHV_OVR
Overload Shutdown activated
3
T_OVR
Overtemp Shutdown activated
4
SUM_SDN
SUM Shutdown: Shutdown latch set
5
SDN_RST
SUM Shutdown Reset signal active
6
DIS_SDNL
Shutdown Latch deactivated
7
HV_ON
High voltage active
8
HV_ON_M
HV-ON switch S1 active
9
NVASL
Analog Input (Setpoint) selected
10 NVEN
Setpoint
enabled
11
U4_ERR
Pos HV Error
12
U5_ERR
Neg HV Error
The status bits AXPWFAIL, IL, IHV_OVR, T_OVR, SUM_SDN, U4_ERR, U5_ERR are set when the
corresponding shutdown condition is present or has led to the shutdown.