RS - 

 
0…32767 

Read Status 
Status word (see below) 

RT - 

 
0…100; 
0…100; 
0…100 

Read Temperatures 
VT_PA1 [0.1°C]:  Temperature of output stage pos. branch 
VT_PA2 [0.1°C]:  Temperature of output stage neg. branch 
VT_DC [0.1°C]: 

Temperature of DC/DC converter 

RV - 

 
-3000…3000 

Read Output Voltage  
VOAV [V]:  

Actual value of output voltage 

SD 

0,1 

Select Setpoint Source 
0: analog setpoint VONVA (default) 
1: digital setpoint VONVD 

SQ - 

Shutdown 

Request 

Set Shutdown Latch 

SR - 

Shutdown 

Reset 

Reset Shutdown Latch 

SS 

0,1 

Select Analog Input Source Resistance 
0: 0

Ω

 signal source output impedance 

1: 50

Ω

 signal source output impedance (default) 

SV -3000… 

3000 

 

Set Output Voltage Setpoint  
VONVD [10mV]: digital setpoint; resolution 10mV  

 
Status word: 
 

Bit #  Status bit 

Function 

AXPWFAIL 

Aux Power Fail: Failure of auxiliary su15V, -15V, +5.2V 

IL  

Interlock loop open 

IHV_OVR 

Overload Shutdown activated 

T_OVR 

Overtemp Shutdown activated 

SUM_SDN 

SUM Shutdown: Shutdown latch set 

SDN_RST 

SUM Shutdown Reset signal active 

DIS_SDNL 

Shutdown Latch deactivated 

HV_ON 

High voltage active 

HV_ON_M 

HV-ON switch S1 active 

NVASL 

Analog Input (Setpoint) selected 

10 NVEN 

Setpoint 

enabled 

11 

U4_ERR 

Pos HV Error 

12 

U5_ERR 

Neg HV Error 

 
The status bits AXPWFAIL, IL, IHV_OVR, T_OVR, SUM_SDN, U4_ERR, U5_ERR are set when the 
corresponding shutdown condition is present or has led to the shutdown. 
 

Summary of Contents for HA51U-3B2

Page 1: ...1 hivolt de GmbH Co KG Oehleckerring 40 D 22419 Hamburg Germany 49 40 537122 0 49 40 537122 99 info hivolt de www hivolt de HA51U 3B2 High Voltage Amplifier 3000V Operating Manual...

Page 2: ...eck After Delivery 9 4 2 Warning Notices 9 4 3 Temperature Compensation 9 4 4 Installation 10 4 5 Power Supply 10 4 6 Ventilation 10 4 7 Functional Test 11 4 8 Input Connection 11 4 9 Analog Control X...

Page 3: ...function the device should immediately be put out of service and it should be secured against unintentional or accidental operation To provide protection for personnel in case of unit failure the safe...

Page 4: ...protected against overload short circuit overtemperature transient overvoltage and high voltage flashover The LEDs on the front panel indicate operation and error statuses An isolated USB interface i...

Page 5: ...functions The signal input is connected to a differential amplifier to avoid interference from ground loops It is available on SMB connector X2 and on X1 Analog Control It provides the analog setpoin...

Page 6: ...pends on the load capacitance The effective load capacitance CL consists of the amplifier s internal output capacitance ca 300pF capacitance of the output line a typical coaxial cable ca 100pF m and c...

Page 7: ...h voltage sources will be switched on The device will then still be in the Interlock shutdown state The high voltage generation is not functionally activated till the Interlock shutdown state is finis...

Page 8: ...ansportation 20 25 50 70 C C Relative humidity Operation Storage and Transportation Not condensing 5 5 80 95 Depending on the ambient temperature modulation amplitude and load capacitance the maximum...

Page 9: ...against any contact before the device is put into operation The test set up must be checked each time before the device is put into operation to ensure that it is not potentially dangerous It should...

Page 10: ...connection is made via the X4 Power Interlock connector at the 24V In and GND pins Caution Before you turn on the voltage source please check its output voltage Admissible range 24VDC 10 If the amplif...

Page 11: ...itch to HV on 7 The Power Enable and HV on LEDs will light up the other LEDs will not 8 Set the switch to HV off 9 The HV on LED will turn off 10 Disconnect the supply voltage 11 Open the interlock br...

Page 12: ...3 I Digital input TTL level 12 GPI4 I Digital input TTL level 13 5VX O Auxiliary voltage output 5V max 10mA 14 GND Supply voltage GND 15 GND Supply voltage GND 16 IL_R O Interlock bridge to X1 IL_IN 1...

Page 13: ...er no software download internal program update is possible 4 11 Power Interlock X4 Connector type Mini Combicon 5 pin Mating connector Phoenix Mini Combicon 5 pin FK MCP1 5_ 5 ST 3 81 Pin Signal Dire...

Page 14: ...N S3 7 Must always be OFF OFF S3 8 Disable X1 Interlock ON Disable ON 4 15 LEDs LED Function Power on 24V supply voltage Overload on overload blink shutdown due to overload Overtemp on overtemperature...

Page 15: ...functions The interlock circuit must be closed and an analog setpoint must be supplied via X1 or X2 High voltage is released by S1 Configuration control shutdown states reset and measuring functions...

Page 16: ...N Parameter par par optional depending on the command Number format 1234 56 leading zeros decimal point and decimal places optional Parameter separator err 1N Error number cr End of record Carriage Re...

Page 17: ...voltage IHVPAV A Actual value of positive HV supply current IHVNAV A Actual value of negative HV supply current VDC mV Actual value of supply HV DC DC converter 5V mV Actual value of 5 2V supply 15V...

Page 18: ...Resistance 0 0 signal source output impedance 1 50 signal source output impedance default SV 3000 3000 Set Output Voltage Setpoint VONVD 10mV digital setpoint resolution 10mV Status word Bit Status b...

Page 19: ...manually In the Device Manager entries are displayed under USB Controller USB Serial Converter and under Ports COM LPT USB Serial Port COMxx The baud rate is set to 115 kBd It cannot be changed Format...

Page 20: ...not light up No supply voltage present HV on display will not light up S1 switch is in HV off position Interlock circuit is open X1 ENABLE signal not set to H Enabling by command EN is switched off O...

Page 21: ...s Low Voltage Directive 2014 35 EU EMC Directive 2014 30 EU and comply with the following European standards EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and la...

Reviews: