10
hivolt.de
GmbH & Co. KG
Oehleckerring
40
∙
D-22419
Hamburg
∙
Germany
∙
+49 40 537122-0
∙
+49 40 537122-99
∙
∙
www.hivolt.de
4.4
Installation
The device can be used as a tabletop unit or installed in a device housing.
At the bottom of the housing there are six M3 threaded holes for the in-housing installation. The
maximum screw-in depth is
4mm!
When installing the device, please remove the rubber feet.
4.5
Power Supply
The device requires an external DC voltage source of 24V
DC
/ min. 1.6A as the power supply. The
connection is made via the
X4 Power / Interlock
connector at the
+24V In
and
GND
pins.
Caution:
Before you turn on the voltage source, please check its output voltage. Admissible range:
+24V
DC
±10%. If the amplifier is operated with wrong supply voltage, it could become damaged.
4.6
Ventilation
Sufficient ventilation is required to remove the heat dissipated inside the case during operation (P
D
of
up to 36W). For tabletop operation, free flow of air is required. Active ventilation could be useful,
depending on the working point and ambient temperature.
If built into a device housing, the thermal losses can be removed through the structure of the housing,