-43-
MN1-5531 Rev. 10
5-12-3. Liquid sterilization (USA only)
•
Applicable chemical solution for sterilization
The applicable sterilants are listed below.
General name
Trade name
Manufacturer
Hydrogen peroxide
PERASAFE
®
*
Practical liquid 1.62W/V%
ANTEC INTERNATIONAL
Peracetic acid
Acecide
®
*
Solution 6%
Saraya Co., Ltd
.
Glutaraldehyde
WAVICIDE
®
-01 *
Solution 2.65%
Medical Chemical Corporation
Glutaraldehyde
STERANIOS *
Solution 2.0%
Laboratoires ANIOS
Glutaral
Cidex plus
®
Solution 3.4%
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Note: * indicates that the marked sterilant is not applicable in Canada.
Warning
After chemical sterilization, thoroughly rinse the probe with sterile water.
Residual sterilant can cause an adverse reaction to the operator or patient.
5-12-4. STERIS
®
sterilization
The applicable product is listed below.
General name
Trade name
Manufacturer
Peracetic acid
STERIS SYSTEM 1E
®
STERIS
®
Regarding the operation of the sterilizer, refer to the documentation supplied with the sterilizer.
Caution
Be sure to attach the waterproof cover (MP-2790).
The connector cannot be soaked in liquid without the waterproof cover.
5-13. Storage
For details about the storage environmental conditions, see section 2-5-2 “Storage environmental conditions”.
Summary of Contents for UST-9120
Page 2: ...MN1 5531 Rev 10 ii ...
Page 15: ... 9 MN1 5531 Rev 10 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 5531 Rev 10 ...
Page 24: ... 18 MN1 5531 Rev 10 ...
Page 50: ... 44 MN1 5531 Rev 10 ...
Page 52: ... 46 MN1 5531 Rev 10 ...
Page 54: ... 48 MN1 5531 Rev 10 ...
Page 58: ... 52 MN1 5531 Rev 10 ...