5-7. Automated cleaning and disinfecting
Probe (with waterproof cover)
The following items must be provided prior to automated cleaning and disinfection:
a) Washer disinfector: according to ISO 15883 with chemo-thermal program (temperature: max 60 °C [140 °F]).
b) Waterproof cover for probe connector MP-2790
c) Detergent: Korsolex Endo-Cleaner (Bode Chemie; # 972 020)
d) Disinfectant: Korsolex Endo-Disinfectant (Bode Chemie; # 972 030)
1) The parameters of the cleaning and disinfection of the device are as follows:
Program step
Water
(40l)
Dosage
(ml/l)
Temp.
(°C)/(°F)
time
(min)
Pre-Rinse
Cold water
5
Cleaning
Deionized water
5 (0.5%)
50/122
10
Rinse
Deionized water
1
Disinfection
Deionized water
10 (1%)
55/131
5
Rinse
Deionized water
1
Rinse
Deionized water
55/131
1
Drying
55/131
15
2) Connect the waterproof cover MP-2790 to the probe connector and confirm there is no air leak by the
tester. About detail information, refer to the instruction manual of the waterproof cover MP-2790.
3) After closing the door, start the chemo-thermal program
4) Open the door after the end of the program.
5) Check whether the probe is dry. If not, proceed as described under drying The applicable chemical
solutions are listed below.
Summary of Contents for UST-9120
Page 2: ...MN1 5531 Rev 10 ii ...
Page 15: ... 9 MN1 5531 Rev 10 Label A 2 Storage case Label B ...
Page 22: ... 16 MN1 5531 Rev 10 ...
Page 24: ... 18 MN1 5531 Rev 10 ...
Page 50: ... 44 MN1 5531 Rev 10 ...
Page 52: ... 46 MN1 5531 Rev 10 ...
Page 54: ... 48 MN1 5531 Rev 10 ...
Page 58: ... 52 MN1 5531 Rev 10 ...