-20-
Q1E-EP1011
B)
Magnetic sensor attachment and Spacer for EZU-RV2S
1)
Prepare the disinfectant solution as stated in the procedure for the
probe.
2)
Immerse the magnetic sensor attachment and spacer for EZU-RV2S into
the disinfectant (see Fig. 13). Set a clock to insure the recommended
contact time which is 5 minutes.
3)
Rinse the magnetic sensor attachment and spacer for EZU-RV2S with
deionized water for 1 minute. (alternatively: immerse it in a tray
filled with deionized water (see Fig.13) for 5 min.)
4)
Visually check its outer surface for leavings of the disinfectant. If
necessary, repeat the rinsing.
5.4
Drying
1)
Wipe the probe with a single-use, fluff-free wipe or towel to remove
moisture from the surface of the probe.
2)
Dry the probe naturally in an ambient temperature between 15-30°C for
a minimum of 4 hours. Alternatively the equipment can be dried using
a drying heater at a temperature of less than 60°C.
Fig.13 Immersion of the probe, the Magnetic sensor
attachment and the Spacer for EZU-RV2S
Water
Detergent
Disinfectant
Drying