-17-
Q1E-EP1011
5.3
Manual Cleaning and disinfection
Prepare following items before manual
cleaning and disinfection:
a)
Detergent: Cidezyme (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device
b)
Disinfectant: Cidex OPA (Johnson & Johnson,
# 20391) or another disinfectant with
approved material compatibility for this
medical device
c)
Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient size
for immersion of the immersible part of the probe at full length)
d)
Soft, fluff free cloth or single use towel
e)
Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A)
EUP-C715
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer’s instruction. If a differing detergent
is used, please also note the approved material compatibility for the
medical device.
2)
Immerse the immersible part of the probe without connector into the
diluted detergent solution (see Fig.13). Wipe the immersible part of
the probe under the surface of the detergent solution with a soft cloth
Manual Cleaning
Rinsing after manual
cleaning
Manual Disinfection
Rinsing after manual
disinfection
Manual cleaning and
disinfection