CAUTION
5)
Use a sterile probe cover to protect the probe. The probe cover
should be allergy free material to avoid allergic reaction.
Between the probe and the probe cover, acoustic coupling gel is
required as a couplant.
6)
Place the probe on the examination site and adjust the probe’s
position for a clear view of the desired image.
7)
After using the probe, the Magnetic sensor attachment and the
Spacer for EZU-RV2S, perform the reprocessing procedure in
accordance with the procedure stated in “
5. Reprocessing procedure
”
every time immediately after completing the ultrasound
examination.
8)
Store the probe, the Magnetic sensor attachment and the Spacer
for EZU-RV2S in the environment indicated in “
6.Maintenance and Safety
inspection
”.
Do not use the probe if the image and the frequency do not match the
probe. An incorrect acoustic output can result in burns or other
injuries to the patient.
Scan for the minimum length of time necessary for the diagnosis and
at the lowest suitable output. There is the possibility that the
patient’s tissues could be affected. For details about the acoustic
output, please refer to the operation manual of the ultrasound
diagnostic instrument.