-23-
Q1E-EP0416
the detergent solution with a soft cloth to remove all visible soil.
3)
The Magnetic sensor attachment and the Spacer for EZU-RV2S should
be left in the detergent solution according to the specified contact
time of the detergent manufacturer.
4)
Rinse the Magnetic sensor attachment and the Spacer for EZU-RV2S with
running tap water for 1 minute. (alternatively: immerse them in a
tray filled with deionized water/tap water (see Fig.15) for 5 min.)
5)
Visually check the outer surface of the Magnetic sensor attachment
and the Spacer for EZU-RV2S for cleanness. If necessary, use
magnifying glass for visually check. If there is still soil visible,
repeat all above steps.
Manual disinfection:
A)
EUP-C532
1)
Prepare the disinfectant solution in a tank with cold water (please
follow the instructions of the disinfectant manufacturer regarding
application, concentration, microbiological efficiency, service
life and contact time).
2)
Confirm the concentration of the disinfectant before immersing the
probe. Although Cidex® OPA does not need to be diluted, it is
recommended to use test strips to verify the concentration. The test
strips can indicate whether or not the concentration is above the
Minimum Effective Concentration (MEC). Please also note the
expiration date of the test stripes. Temperature of disinfectant
solution should be minimum 20 °C. The minimum contact time is 5 minutes.
If a different disinfectant is used, follow the manufacturer’s
instructions. Please also consider the material compatibility for the
medical device.
3)
Immerse the immersible part of the probe into the disinfectant (see
Fig. 15). Set a clock to insure the recommended contact time which
is 5 minutes.
4)
Rinse the immersible part of the probe with deionized water for 1 minute.
(alternatively: immerse the immersible part of the probe in a tray
filled with deionized water (see Fig.15) for 5 min.)
5)
Visually check the outer surface of the probe for leavings of the
disinfectant. If necessary, repeat the rinsing.