-24-
Q1E-EP0416
B)
Magnetic sensor attachment and Spacer for EZU-RV2S
1)
Prepare the disinfectant solution as stated in the procedure for the
probe.
2)
Immerse the Magnetic sensor attachment and the Spacer for EZU-RV2S
into the disinfectant (see Fig.15). Set a clock to insure the
recommended contact time which is 5 minutes.
3)
Rinse the Magnetic sensor attachment and the Spacer for EZU-RV2S with
deionized water for 1 minute. (alternatively: immerse them in a tray
filled with deionized water (see Fig.15) for 5 min.)
4)
Visually check the outer surface of the Magnetic sensor attachment
and the Spacer for EZU-RV2S for leavings of the disinfectant. If
necessary, repeat the rinsing.
5.4
Drying
1)
Wipe the equipment with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the equipment.
2)
Dry the probe naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the probe can be dried using
a drying heater at a temperature of less than 60°C.
Fig.15 Immersion of the probe, the Magnetic sensor attachment
and the Spacer for EZU-RV2S
Water
Detergent
Disinfectant
Drying