-35-
Q1E-EP1457
WARNING
5.5
Drying
1)
Wipe all parts of the needle guide bracket with a single-use,
fluff-free wipe or towel to remove moisture from the surface of them.
2)
Dry all parts of the needle guide bracket naturally in an ambient
temperature between 15-30°C for a minimum of 4 hours. Alternatively
they can be dried using a drying heater at a temperature of less
than 55°C.
5.6
Packaging
Packing the device into the sterile bag is recommended until using it
for keeping cleaned and disinfected condition.
5.7
Assembly and Inspection
Assemble the needle guide bracket and inspect it for any damage such
as crack, scratch or deformation. Do not use it if any damage is found.
5.8
Storage
Store the equipment in a cool, dustproof and dark space to avoid
high temperature, humidity and direct sunlight.
Do not perform any sterilization procedure to the needle guide bracket.
The needle guide bracket is not applied to any sterilization procedure.
If sterilization procedure is performed to it, it may suffer serious
damage and become unusable.
6.
Maintenance and Safety Inspection
1)
After using the probe and the attachment, they should be cleaned
and disinfected or sterilized according to “4. Reprocessing
Procedure”, then store them in a cool and dark place avoid high
temperature and humidity, direct sunlight.
2)
After using the Needle guide bracket, it should be cleaned and high
level disinfected according to “4. Reprocessing Procedure”, then
store it in a cool and dark place avoid high temperature and humidity,
direct sunlight.
3)
Visually inspect the surface of the probe head, housing, cable
connector and the attachment for any crack, scratch or
denaturalization. If you find any damage, do not use the probe and
contact a service support immediately.
4)
Visually inspect the surface of the Needle guide bracket for any
crack, deformation or denaturalization. If you find any damage, do
not use it and contact a service support immediately.
Drying
Summary of Contents for C22P
Page 48: ... 42 Q1E EP1457 Unit mm Fig 23 Dimensions ...
Page 49: ... 43 Q1E EP1457 Unit mm Fig 24 External view with the attachment ...
Page 50: ... 44 Q1E EP1457 Unit mm Fig 25 Dimensions with the Needle Guide Bracket ...
Page 51: ... 45 E Q1E EP1457 Fig 26 External view with Ultra Pro II Needle Guide cover kit Unit mm ...