-34-
Q1E-EP1457
5.4
Automated cleaning and disinfection
The following items must be provided prior to automated cleaning and
disinfection:
a)
Washer disinfector: according to DIN EN ISO 15883 with
chemo-thermal program (temperature: max. 60°C)
b)
Detergent: Korsolex® Endo-Cleaner (Bode Chemie; #972020) or
another cleaning agent with approved material compatibility for
this medical device
c)
Disinfectant: Korsolex® Endo-Disinfectant (Bode Chemie;
#972030) or another disinfectant with approved material
compatibility for this medical device
d)
Washer disinfector accessories
basket for holding the needle guide bracket
basket with lid for holding the dismantled small parts (screws,
spring) of the needle guide bracket
1)
The parameter of the cleaning and disinfection of the device are
as follows:
Program step water (40 l)
Dosage (ml/l) Temp. (°C) time (min)
Pre- Rinse
Cold water
5
Cleaning Deionized water
5 (0,5 %)
50
10
Rinse Deionized
water
1
Disinfection Deionized water
10 (1 %)
55
5
Rinse Deionized
water
1
Rinse Deionized
water
55
1
Drying
55
15
Table 2: Parameters for automated process
2)
Place the pre-cleaned and dismantled needle guide bracket and
the dismantled parts (srews, spring) into the basket. Small parts
must be placed in a strainer basket with lid.
3)
After closing the door, start the chemo-thermal program.
4)
After end of the program, open the door.
5)
Remove the dismantled needle guide bracket and check whether it
is dry. If not, proceed as described under drying.
Summary of Contents for C22P
Page 48: ... 42 Q1E EP1457 Unit mm Fig 23 Dimensions ...
Page 49: ... 43 Q1E EP1457 Unit mm Fig 24 External view with the attachment ...
Page 50: ... 44 Q1E EP1457 Unit mm Fig 25 Dimensions with the Needle Guide Bracket ...
Page 51: ... 45 E Q1E EP1457 Fig 26 External view with Ultra Pro II Needle Guide cover kit Unit mm ...