31
The following graph is the relationship of the partial scan and "capture width", "frame rate".
Equation below is the formula for the "total number of lines by capture width (the decimal point is truncated)" and the "frame
rate".
(1) Equation of total number of line for partial scan mode
The total number of line = 33+
OFFSET/12
+
(1207-HEIGHT-OFFSET)/12
(
is Ceiling function )
(2) Equation of total number of line for partial scan and binning mode
The total number of line = 22+
OFFSET/12
+
HEIGHT/2
+
(1202-HEIGHT-OFFSET)/12
(
is Ceiling function )
Frame rate = (45000000 / 1920) / The total number of line
*Notes on partial scan use
The capture start po capture width, please use 1201 or less.
30
40
50
60
70
80
90
100
110
120
130
200 225 250 275 300 350 400 450 500 550 600 700 800 900 1000 1100 1200
partial_binning
(SensorArea)
partial_binning
(OutputLine)
partial
(SensorArea
&OutputLine)
Fr
am
e
Ra
te
[F
PS
]
Vertical Line (SensorArea & OutputLine) [H]