4.2 Timing Chart
37
Timing Chart Interval Descriptions
Interval
Description
Duration
t1
Channel setup time
100
μ
s or more
t2
Channel hold time
200
μ
s or more
t3
OUTPUT ON
→
Busy delay time
200
μ
s or less
t4
OUTPUT ON
→
Voltage output delay time
600
μ
s or less
t5
Alarm delay time
3.5 ms or less
t6
Output pulse width
8 ms or more
t7
OUTPUT OFF
→
Busy delay time
800
μ
s or less
t8
OUTPUT OFF
→
Voltage output delay time
2.5 ms or less
t9
Output off time
4 ms or more
t10
OUTPUT ON
→
Discharging start delay time
1.2 ms or less
t11
OUTPUT OFF
→
Discharging complete delay time
1.1 ms or less