2) output voltage
These tables show the relations between displayed value of SM-8216
super megohmmeter and output voltage of RI-8000.
Target voltage 10V
Displayed value
0.05 0.1 0.5
Output voltage
10V 5V 1V
Target voltage 25V
Displayed value
0.125 0.25 1.25
Output voltage
10V 5V 1V
Target voltage 50V
Displayed value
0.25 0.5 2.5
Output voltage
10V 5V 1V
Target voltage 100V
Displayed value
0.5 1 5
Output voltage
10V 5V 1V
Target voltage 250V
Displayed value
1.25 2.5 12.5
Output voltage
10V 5V 1V
Target voltage 500V
Displayed value
2.5 5 25
Output voltage
10V 5V 1V
Target voltage 1000V
Displayed value
5 10 50
Output voltage
10V 5V 1V
Example
・
SM-8216
Target
voltage : 100V
RANGE
:
×
10
2
Measurement range : 0.5
×
10
2
M
Ω
to 20
×
10
2
M
Ω
・
Output voltage of RI-8000
Output range : 0.5
×
10
2
M
Ω
to 5
×
10
2
M
Ω
Output voltage : 10V/0.5
×
10
2
M
Ω
, 1V/5
×
10
2
M
Ω