27
Target voltage 500V
Displayed value
2.5 12.5 25
Output voltage
1V 5V 10V
Target voltage 1000V
Displayed value
5 25 50
Output voltage
1V 5V 10V
Example
・
SM-8216
Target voltage : 500V
RANGE
:
×
10
4
Measurement range : 2.5
×
10
4
M
Ω
to 100
×
10
4
M
Ω
・
Output voltage of RP-8000
Output range : 2.5
×
10
4
M
Ω
to 25
×
10
4
M
Ω
Output voltage : 1V/2.5
×
10
4
M
Ω
, 10V/25
×
10
4
M
Ω
7.1.2 DC Output, RI-8000
– Inversely proportional or 1/R to resistance
In insulation resistance, as described in 7.1.1 DC Output, RP-8000, if the
measuring voltage is constant, the current flowing through the sample is
halved, and the output voltage is also halved when the insulation
resistance -- R -- is doubled. This relation can be expressed as “1/R.”
The RI-8000 provides this type of DC signal output.
1) Specifications
Output Range: Full range of the measuring range of the SM-8216
super megohmmeter
Output Voltage: 10V/minimum value of each range of
super megohmmeter , 1 V/10 times minimum value
Output Accuracy: Within 10% of displayed value in the range from
minimum value to 10 times minimum value
2) Usage
Measure the insulation resistance of a sample with the method
designed in 6.1 Measuring Method. There is no limitation in the
applications in relation with the provision of the RI-8000 DC output.
Fig. 7.1.1 on the previous page shows the DC OUT (OPTION) connector
on the rear of the unit.
A DC measuring instrument like a HIOKI chart recorder whose input
impedance is greater than 1 M
Ω
can be connected to these terminals.
For permanent data recording, a HIOKI chart recorder is recommended
for better technical follow-up, including a supply of a variety of
recording charts and technical service.