Troubleshooting
Error Displays and Remedies
When an error is displayed on the LCD screen, repair is necessary.
Please contact your authorized Hioki distributor or reseller.
Display
Description
Remedy and Reference
The instrument was
unable to perform
zero adjustment.
(Low resistance
measurement)
•
Verify that there is no broken
connection in the test leads.
•
Zero adjustment can be
performed for readings of up
to 3
Ω
. Ensure that the wiring
resistance is 3
Ω
or less.
The specified fuse has
not been installed.
Use only the specified type of
The settings data has
been corrupted.
Repair is required.
The pre-adjustment
data has been
corrupted.
The measurement
circuit is broken
Replace the batteries.
If there is no apparent
improvement, the instrument
needs repair.
The voltage generation
circuit is broken.
Bluetooth
®
is broken.
Repair is required.
There is a broken
connection in the
suprotective fuse.
Replace the indicated fuse.
Low battery voltage