33
Current Measurement
Automatic hold conditions
Displayed value will freeze when the following two conditions are satisfied at the same time:
•
When the measured value fluctuation stabilizes within the fluctuation range described in the table on the
next page.
•
When the measured value exceeds the threshold value described in the table on the next page (voltage,
current).
When the measured value is less than the threshold value described in the table on the next page (resistance,
continuity check, diode).
Start
End
Start End
Start
End
Threshold value
e.g., 12.0 A
Automatic hold
e.g., 99.0 A
Automatic hold,
e.g., 100.0 A
Measured
value
Time
: Fluctuation range
: Measured value to be frozen
(1) Not frozen automatically
(Does not fall below threshold.)
(2) Not frozen automatically
(Does not exceed threshold.)
(1)
(2)
Suppose that the measured value has fallen below (voltage, current) or exceeded the threshold value
(resistance, continuity, diode) after the displayed value froze. When the two auto-hold conditions are met
again after that, refreshing measured values will stop.
HIOKI CM4371C961-00