70
Input Specifications, Measurement Specifications
(4)
Other measurement parameters
Overload protection
1000 V DC, 1000 V AC, or 2 × 10
7
V·Hz, whichever is lower
(Up to 1 minute of continuous application)
Overload current
At steady state: 30 mA or less
At transient state: 1.5 A or less
Display update rate*
Capacitance
0.5 to 5 times/s (depending on the capacitance)
Temperature
(Type K thermocouple)
1 time/s (including thermocouple wiring break check)
Response time
Continuity check
Open or short-circuit condition lasting for 0.5 ms or
more can be detected.
Open terminal voltage
Continuity check,
resistance, diode
2.0 V DC or less
Continuity threshold
25
Ω
±10
Ω
(continuous beep, LCD backlighted in red)
Discontinuity
threshold
250
Ω
±10
Ω
Stabilization time for
reference junction
compensation of
instrument
Up to 120 minutes
(Reference: when the instrument having a temperature of 23°C is left to stand in
ambient environments of 65°C for 60 minutes)
*: Does not include range switching time.
HIOKI CM4371C961-00