29
__________________________________________________
Chapter 2 Measurement Procedure
_____________________________________________
MAX
→
MIN
→
AVE
→
instantaneous value
(no indication)
3. During measurement, the
MAX/MIN
key can
be used to select the value that should be
shown.
MAX
: Maximum value is shown.
MIN
: Minimum value is shown.
AVE
: Average value is shown = (maximum
value + minimum value)/2
If none of the
MAX
,
MIN
, or
AVE
indicators is
shown, the display shows the instantaneous
value.
4. Pressing the
HOLD
key will stop the recording
function. The
HOLD
indication appears and the
REC
indication stops flashing.
By pressing the
MAX/MIN
key in this
condition, the
MAX
,
MIN
, and
AVE
values
stored in the internal memory can be called up
on the display, as follows.
While
HOLD
is shown, the elapsed time is not
incremented. Also if the clamp sensor is
removed from the conductor for easier reading,
the minimum value will not return to zero.
Pressing the
HOLD
key once more causes the
HOLD
indication to go out. The recording
function resumes, and
REC
flashes again.