52
────────────────────────────────────────────────────
7.4 Commands Specific to the 3157
────────────────────────────────────────────────────
:SYSTem:OPTion:BUZZer
■
Sets the buzzer.
:SYSTem:OPTion:BUZZer?
■
Queries the buzzer.
Syntax
:SYSTem:OPTion:BUZZer
_<data>
<data>
Numerical data in NR1 format between 0 and 3
Function
Sets the buzzer.
1: ON at screening, ON at error
2: OFF at screening, OFF at error
3: OFF at screening, ON at error
4: ON at screening, OFF at error
The numerical value can be in NRf format, but any digits after the decimal
point will be rounded.
Example
Transmission
:SYSTem:OPTion:BUZZer 3
The buzzer is set to ON at screening, OFF at error.
Error
If <data> is other than numerical value described above, an execution error
occurs.
The execution of this command in a state other than the following states
causes an execution error:
・
READY state
・
Optional function setting screen
Syntax
:SYSTem:OPTion:BUZZer?
Function
Returns the buzzer setting as a numerical value in NR1 format.
<data>
Numerical data in NR1 format between 0 and 3
Example
Response
If headers are on
:SYSTEM:OPTION:BUZZER 3
If headers are off
3
Error
If the response message is longer than 300 bytes, a query error is generated.