85
────────────────────────────────────────────────────
4.14 Setting and Activating the Comparator
────────────────────────────────────────────────────
Shows the results of the decisions
made by the comparator.
Shows the first decision
condition setting.
Shows the third decision
condition setting.
NOTE
4.
Set the first and third parameters to be the two parameters which you wish
to measure. (Refer to the section on display parameter setting for details
upon how to make the setting.)
The Conditions setting screen is accessed by pressing the
key.
5.
The Menu screen can be displayed from the Comparator screen display by
pressing the
key, in the same manner as from the Initial screen, and
then the values for the various test conditions can be changed. (For details,
refer to the explanation of setting the appropriate test condition.)
When the trigger is internal, testing starts at the time point at which the
comparator mode is started, and the results of the decisions will be output
through the EXT I/O terminal.
The test conditions during comparator operation and normal operation are
compatible.