39
────────────────────────────────────────────────────
4.4 Setting the Test Signal Level
────────────────────────────────────────────────────
4.4.1 Control Screen Sequence
Menu screen
Test signal level setting screen
Set open
circuit voltage
(V)
Set constant
voltage (CV)
Set constant
current (CC)
Initial screen
Initial screen
4.4 Setting the Test Signal Level
The value of the test signal level may change according to the sample which
is being tested. With the 3522-50, it is possible to vary the level of the test
signal applied to the object under test over a wide range using the following
three methods:
Set open circuit voltage (V)
The value of the open circuit voltage is set.
Set constant voltage (CV)
The value of the voltage between the
terminals of the object under test is set.
Set constant current (CC)
The value of the current flowing through the
object under test is set.
Press the
key
Press the
key
Select the type of level for the test signal (open
circuit voltage, constant voltage, or constant
current) to be set.
Input the value for the test signal level.
When the setting process is completed, press the
key to return to the Initial screen.