30
────────────────────────────────────────────────────
4.1 Description of the Screens
────────────────────────────────────────────────────
4.1.1 The Initial Screen
2
1
3
4
5
6
7
8
9
10
11
12
13
14
1. Measured values
2. Parameter keys
(The parameters displayed can be
selected using these keys.)
3. FREQ
Test frequency
4. V
Test signal control level
5. Vmoni
Voltage value between the
terminals of the test sample
6. I-LIM
Measurement current limit value
7. Imoni
Current value flowing through
the test sample
8. RANGE
Test range
9. OPEN
Open circuit compensation setup
10. SHORT
Short circuit compensation setup
11. TRIG
Trigger mode
12. DELAY
Set value for the trigger delay
13. AVE
Number of times for averaging
14. SPEED
Testing speed
NOTE
This means that the 3522-50 is operating
outside the range for which the accuracy
of the measured values can be assured.
The Initial screen is the screen which is initially displayed when the power
to the 3522-50 unit is turned on.
The monitor display on this screen enables the user to perform testing while
checking the values of all the test conditions. (The monitor display varies
according to the test signal level setting.)
Even if the 3522-50 unit is being used outside the limits of its specification,
test values are sometimes displayed. In this case, the following type of
indication appears on the display.
In this case, you should consider the following possible causes, and you
should either change the test conditions while checking the accuracy assured
ranges (with reference to Section 7.5, "Test Accuracy"), or you should
consider the measured values as values for reference only, because their
validity is somewhat doubtful.
Perhaps the test signal controlled level is too low.
Increase the test signal level.
Perhaps the present test range (if ranging is set to HOLD) is not suitable.
Either set the test range to the most suitable one by using AUTO ranging, or
change the test range manually.
When powering on again, the screen when the power has been turned off on
the following screens is displayed.
Initial screen, Comparator screen, Scaling screen, Magnification screen,
Continuous test screen
Summary of Contents for 3522-50
Page 1: ...Instruction Manual For 3522 50 LCR HiTESTER June 2011 Revised edition 7 3522C981 07 11 06H...
Page 2: ......
Page 8: ......
Page 20: ...6 1 3 Names and Functions of Parts...
Page 170: ...156 5 9 9442 PRINTER option...
Page 197: ...183 7 5 Test Accuracy Conversion Table from C and L to Z Frequency Hz...
Page 198: ...184 7 5 Test Accuracy...
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......