2.5 Auto-Zero-Adjustment/
Zero-Cancel Correction Function
2.5.1 Auto-Zero-Adjustment Function
The auto-zero-adjustment function is used to adjust
offsets in the internal circuit automatically that
result from temperature characteristics or clamp
sensor magnetization. The clamp core is
magnetized during a large DC current measurement,
or when a powerful magnet is placed close to the
clamp core.
1. Wait until the counter is stable under no input.
Then, press
0ADJ/RESET
.
ADJ
annunciator lights.
・
You can perform the auto-zero-adjustment, if the
counter displays within
4.5 A in a current mode.
・
When there is an input or the counter decreases, the
measurement accuracy will be spoiled by pressing
0ADJ/RESET as well as the accurate auto-zero-
adjustment. If inaccurate auto-zero-adjustment is
performed, perform the correct procedure again.
・
It would take approximately 20 seconds to stabilize
the counter in AC+DC A mode.
・
Use the zero-cancel correction function if the
counter fails to revert to zero after correct auto-zero-
adjustment in AC+DC A mode.
・
If you press
0ADJ/RESET
again during the auto-
zero-adjustment in the internal circuit, the auto-zero-
adjustment is canceled.
・
Do not perform auto-zero adjustment when using
auto range in the DC V mode. Always switch to the
range you will use (a manual range) before
performing the adjustment.
Summary of Contents for 3284 HiTester
Page 2: ......
Page 17: ...13 第1 章 製品概要 3284 1 3 各部の名称と機能 ...
Page 35: ...31 第2 章 測定方法 20 A 10 A 図2 出力応答波形 立上り 250 ms 出力波形 入力波形 500 ms 図3 出力応答波形 立下り 出力波形 入力波形 ...
Page 58: ...54 第4 章 電池の交換方法 ...
Page 60: ...56 第5 章 AC アダプタ 別売 の使用 ...
Page 61: ...57 第6 章 ハンドストラップの付け方 第 6 章 ハンドストラップの付け方 ハンドストラップを付けると操作性が増します ...
Page 62: ...58 第6 章 ハンドストラップの付け方 ...
Page 68: ...64 第7 章 故障とお考えになる前に ...
Page 70: ...66 第8 章 アフターサービス ...
Page 71: ...3284 ...
Page 72: ......
Page 73: ...3284 CLAMP ON AC DC HiTESTER INSTRUCTION MANUAL ...
Page 74: ......
Page 93: ...3 Chapter 1 Product Outline Top and Side View 1 3 Parts and Functions ...
Page 132: ...42 Chapter 2 Measurement Procedure ...
Page 144: ...54 Chapter 4 Battery Replacement ...
Page 148: ...58 Chapter 6 Attaching The Hand Strap ...
Page 156: ...66 Chapter 8 Service ...
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......